- Stainless steel DIN 1.4404
- Maximum flow rate: 1000 ml/min
- Maximum operating pressure: 400 bar
- Temperature range: RT – 100 °C
- Dead-Volumen: ca. 27 ml
- Heating Power: 3600 W
Device for precise and stable temperature control of eluents for preparative chromatography
Optional:
- column heating jacket
- 2 x External Temperature measurement (PT100)
Temperature settings and display::
Manually via touch screen
Via network or RS-232
Via software: PrepCon or SCPA Templogger
Sensors: Pt100
Control accuracy: typical +0.5%
Calibration of all temperature sensors and heating controllers by the end user possible
Flowrate | Maximum Δ T (Water) | Backpressure |
500 ml/min | 60 K | 6 bar |
700 ml/min | 50 K | 11 bar |
1000 ml/min | 40 K | 20 bar |
Temperature control range | Room temperature (RT) bis 100°C |
Heating power | 3600 W |
Heating distance | 8 Meter Stainless steel capillary |
Flowrate | up to approx. 1000 ml/min |
Capillaries | 1/8″ with 2.1 mm inner diameter |
Capillary connection | Swagelok® Pipe fitting |
Dead volume | approx. 27 ml |
Wetted materials | Stainless steel DIN 1.4404 | AISI/SAE 316L (others possible on request) |
System protection | Thermal insulation, temperature protection switch (110°C) |
Dimensions (H x B x T) | 18 x 30 x 35,5 cm |
Weight | 22 kg |
Temperature acquisition | Pt100 Resistance thermometer |
Channels | Any combination of up to 8 control and display channels possible, maximum 4 operable via touch screen |
Communication | Ethernet or RS-232 (Serial) |
Power consumption | 3600 W |
Power supply | 230 V AC |